Difference between revisions of "Pt master recipe"

From Quattrone Nanofabrication Facility
Jump to navigation Jump to search
(Pt master recipe)
 
 
(3 intermediate revisions by the same user not shown)
Line 1: Line 1:
[[Denton_Explorer14_Magnetron_Sputterer | Return to PVD-05 ]]
+
> [[Denton_Explorer14_Magnetron_Sputterer#Process_data_&_master_recipes | Click here to return to PVD-05]] <
 +
 
 +
Step T006 is the deposition step.
 +
 
 +
{| class="wikitable"
 +
! Step Number || T000 || T001 || T002 || T003 || T004 || T005 || T006 || T007 || T008 || T009 || T010 || T011 || T012 || T013
 +
|-
 +
|-
 +
| Step Time (sec) || 5 || 30 || 30 || 30 || 30 || 30 || [dep] || 30 || 30 || 30 || 30 || 30 || 400 || 5
 +
|-style="background-color:#FFFFFF"
 +
| Min Vacuum Setpoint (Torr) ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-
 +
|  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  || 
 +
|-style="background-color:#DBFEFF"
 +
| Gas - (PID or Fixed) ||  || PID || PID || PID || PID || PID || PID || PID || PID || PID || PID || PID ||  ||
 +
|-style="background-color:#FFFFFF"
 +
| Gas - PID Master Gas Select ||  || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 ||  ||
 +
|-
 +
| Gas1 - Setpoint (sccm) ||  || 10 || 10 || 10 || 10 || 10 || 10 || 10 || 10 || 10 || 10 || 10 ||  ||
 +
|-style="background-color:#FFFFFF"
 +
| Gas2 - Setpoint (sccm)  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-
 +
|  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#DEB4DC"
 +
|Gas PID Pressure (mTorr) ||  || 3 || 3 || 3 || 3 || 3 || 3 || 3 || 3 || 3 || 3 || 3 ||  ||
 +
|-style="background-color:#DCF5E9"
 +
| RF Source - Sputter (Watts)  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#DCF5E9"
 +
| RF Source - Cathode Select  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#DCF5E9"
 +
| RF Source - Shutter  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#ECFA6F"
 +
| DC 1 Source - Sputter (Watts) ||  || 25 || 50 || 75 || 100 || 125 || 140 || 120 || 100 || 80 || 60 || 25 ||  ||
 +
|-style="background-color:#ECFA6F"
 +
| DC 1 Source - Cathode Select ||  || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 || 1 ||  ||
 +
|-style="background-color:#ECFA6F"
 +
| DC 1 Source - Shutter ||  ||  ||  ||  ||  ||  || Open ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#FAD86F"
 +
| DC 2 Source - Sputter (Watts) ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#FAD86F"
 +
|  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  || 
 +
|-style="background-color:#FAD86F"
 +
| DC 2 Source - Shutter ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#6FD2FA"
 +
| Pressure Control (Throttle) || Yes || Yes || Yes || Yes || Yes || Yes || Yes || Yes || Yes || Yes || Yes || Yes ||  ||
 +
|-style="background-color:#FFFFFF"
 +
| Ignition Pressure (mTorr) ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-
 +
|  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||
 +
|-style="background-color:#E8F213"
 +
| Rotation Speed (0-100%) || 50 || 50 || 50 || 50 || 50 || 50 || 50 || 50 || 50 || 50 || 50 || 50 || 50 ||
 +
|-
 +
| End Process (Yes) ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  ||  || Yes
 +
|}

Latest revision as of 11:32, 31 January 2023

> Click here to return to PVD-05 <

Step T006 is the deposition step.

Step Number T000 T001 T002 T003 T004 T005 T006 T007 T008 T009 T010 T011 T012 T013
Step Time (sec) 5 30 30 30 30 30 [dep] 30 30 30 30 30 400 5
Min Vacuum Setpoint (Torr)
Gas - (PID or Fixed) PID PID PID PID PID PID PID PID PID PID PID
Gas - PID Master Gas Select 1 1 1 1 1 1 1 1 1 1 1
Gas1 - Setpoint (sccm) 10 10 10 10 10 10 10 10 10 10 10
Gas2 - Setpoint (sccm)
Gas PID Pressure (mTorr) 3 3 3 3 3 3 3 3 3 3 3
RF Source - Sputter (Watts)
RF Source - Cathode Select
RF Source - Shutter
DC 1 Source - Sputter (Watts) 25 50 75 100 125 140 120 100 80 60 25
DC 1 Source - Cathode Select 1 1 1 1 1 1 1 1 1 1 1
DC 1 Source - Shutter Open
DC 2 Source - Sputter (Watts)
DC 2 Source - Shutter
Pressure Control (Throttle) Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes Yes
Ignition Pressure (mTorr)
Rotation Speed (0-100%) 50 50 50 50 50 50 50 50 50 50 50 50 50
End Process (Yes) Yes