Difference between revisions of "SUSS MicroTec MA6 Gen3 Mask Aligner"

From Quattrone Nanofabrication Facility
Jump to navigation Jump to search
 
(One intermediate revision by the same user not shown)
Line 13: Line 13:
 
| NEMO_Designation = MA-01
 
| NEMO_Designation = MA-01
 
| Lab_Phone = XXXXX
 
| Lab_Phone = XXXXX
| SOP Link = [https://www.seas.upenn.edu/~nanosop/MA6_SOP.htm SOP]
+
| SOP Link = [https://nemo.nano.upenn.edu/media/tool_documents/ma-01-ma6-mask-aligner/MA-01_SOP_v01.pdf SOP]
 
}}
 
}}
  
 
== Description ==
 
== Description ==
This system is a dual-use mask aligner and wafer-bond aligner. Mask alignment is used for contact and proximity exposure processes. Exposures can be performed with gaps programmable from 10 um to 300 um in 1 um increments. Automatic wedge error compensation (WEC) is used to ensure that the mask and wafer are parallel. The lamp is a 1000 W Hg-Arc lamp. Integrated light level sensing ensures proper exposure doses as the lamp degrades.
+
This system is a mask aligner set up for piece parts to 6" wafers, with first mask, top-side alignment, bottom-side alignment, and flood exposure. Mask alignment is possible for both contact and proximity exposure processes. Exposures can be performed with gaps programmable from 10 um to 300 um in 1 um increments. Automatic wedge error compensation (WEC) is used to ensure that the mask and wafer are parallel. The lamp is a 1000 W Hg-Arc lamp. Integrated light level sensing ensures proper exposure doses as the lamp degrades.
  
 
===== Applications =====
 
===== Applications =====
 
* First mask exposure of resist-coated samples for patterning
 
* First mask exposure of resist-coated samples for patterning
 
* Aligned mask exposure of resist-coated samples for patterning
 
* Aligned mask exposure of resist-coated samples for patterning
* Flood exposure of resist-coated samples for image reversal (AZ5214)
+
* Flood exposure of resist-coated samples for image reversal
 
* Flood exposure of samples for further cross-linking of resist/resin
 
* Flood exposure of samples for further cross-linking of resist/resin
  

Latest revision as of 11:57, 19 November 2024


SUSS MicroTec MA6 Gen3 Mask Aligner
MA-01.jpeg
Tool Name SUSS MicroTec MA6 Gen3 Mask Aligner
Instrument Type Lithography
Staff Manager Ana Cohen
Lab Location Bay 5
Tool Manufacturer SUSS MicroTec
Tool Model MA6 Gen3
NEMO Designation MA-01
Lab Phone XXXXX
SOP Link SOP

Description

This system is a mask aligner set up for piece parts to 6" wafers, with first mask, top-side alignment, bottom-side alignment, and flood exposure. Mask alignment is possible for both contact and proximity exposure processes. Exposures can be performed with gaps programmable from 10 um to 300 um in 1 um increments. Automatic wedge error compensation (WEC) is used to ensure that the mask and wafer are parallel. The lamp is a 1000 W Hg-Arc lamp. Integrated light level sensing ensures proper exposure doses as the lamp degrades.

Applications
  • First mask exposure of resist-coated samples for patterning
  • Aligned mask exposure of resist-coated samples for patterning
  • Flood exposure of resist-coated samples for image reversal
  • Flood exposure of samples for further cross-linking of resist/resin

Processes

Exposure Parameters

The following should help narrow down potential exposure parameters based on resist type and thickness. Information based on the following resist datasheets -- S1800 Series, SPR 220 Series, SU-8 2000-2015, SU-8 2025-2075, SU-8 2100, SU-8 3000.

Resist Thickness (um) Dose (mJ/cm2) Channel
S1813 1.2 150 #2
SPR220 - 3.0 3.0 310 #1
SPR220 - 4.5 5.0 380 #1
SU-8 2005 5.0 - 6.0 105 - 110 #1
SU-8 2050 40 - 80 150 - 215 #1
80 - 110 215 - 240 #1
115 - 160 240 - 260 #1
SU-8 2100 100 - 150 240 - 260 #1
150 - 225 260 - 350 #1
225 - 270 350 - 370 #1
SU-8 3050 40 - 100 150 - 250 #1
Exposure Wavelengths

The software adjusts the lamp intensity controller during exposure based on the channel selected in the recipe.

  • Channel #1 = 365nm (i-line)
  • Channel #2 = 405nm (h-line)

For some applications it can be beneficial to eliminate wavelengths below 365nm using a long pass filter. Contact staff to be trained on setting this filter if you plan to use it. Since higher energy wavelengths are removed, expect an increase in exposure time of approximately 40% to reach the optimum exposure dose.

Resources

SOPs & Troubleshooting
  • SOP
  • Training Video - First Mask
  • If the recipe cannot be changed directly with the recipe button it typically means that the user management software has been logged out. To fix it, do the following:
    • Minimize the main software
    • Open the 'User Management' software on the desktop
    • Login with username:user and no password