Oxford 80 Plus RIE

From Quattrone Nanofabrication Facility
Revision as of 08:26, 3 November 2022 by Azadi (talk | contribs)
Jump to navigation Jump to search


Oxford 80 Plus RIE
DE-04.jpeg
Tool Name Oxford 80 Plus RIE
Instrument Type Etch
Staff Manager Sam Azadi
Lab Location Bay 2
Tool Manufacturer Oxford Instruments
Tool Model 80+
NEMO Designation {{{NEMO_Designation}}}
Lab Phone 215-898-9748
SOP Link QNF SOP

Description

Oxford 80 Plus is an anisotropic dry etch tool. The main use of this tool is to etch SiO2, and SiNx to create hard masks for deep Si etch. However, it etches many other materials, including a variety of resists. The tool is connected to the following gases: Ar, O2, CF4, CHF3, SF6.

The plasma is ignited by the oscillating electric field created by the RF power at 13.56 MHz. The wafer is cooled down via passive cooling through a graphite chuck.

Applications
  • Etch of SiO2
  • Etch of SiNx
  • Etch of Si
  • Etch of resist


Standard Process Information =

Protocols and Reports


Etch Rate

Recipe parameters are defined in the reports above.

  • Etch rate of PECVD oxide and nitride are listed in the reports above.
  • Etch rate of thermal SiO2 using CF4 recipe: 48 nm/min
  • Etch rate of thermal SiO2 using CHF3/O2 recipe: 45 nm/min
  • Etch rate of LPCVD SiNx using CF4 recipe: 50 nm/min
  • Etch rate of LPCVD SiNx using CHF3/O2 recipe: 57 nm/min


Resources

Oxford 80+ Training Video


SOPs & Troubleshooting